الوصف
Radiation source halogen lamp, adjusted
Beam path irradiation principle
Spectral measuring range UV/VIS range 340, 546, 578 nm interference filter
Insert filter selection individually, coded
Operating modes/methods extinction, concentration with factor, standard, non-linear reference curve, kinetics coagulation according to fixed absorbance
Possible determinations clinical chem. Determinations UV-VIS range HbAl, Hct, glucose in 2 min. , EIA, coagulation
Data entry programmed, access via keypad
Selection of the measuring function buttons
Display factor input digitally - factors already included
Display filter value on the filter, coded
display absorbance Linearity 4 digits, digital and printer output
Display result 4 digits, digital and printer output
Measuring range absorbance: -0.999 . . . 2.600
Measuring range concentration 0.000 . . . 9999
Short-term reproducibility <0.005 E at E = 1.0
photometric Accuracy <1.0% at E = 1.0
False light <0.005%
Long-term stability <0.010 U/h
Minimum measuring volume 450 µl micro; 1.5ml macro
Zero point adjustment: automatic
Substrate calculator: available
Rate/Timer: Available
Cuvette temperature control water circulation thermostat (not included)
Cuvette holder 10 mm rectangle macro, semi-micro, 10 mm suction cuvette
Maximum layer thickness: 10 mm
Measuring time/sample throughput substrates: 200 analyzes/h
Enzymes: 150 analyses/h
Power supply 220 volts, 50 Hz
Power consumption 85VA
Dimensions W-H-D (mm) 425 x 175 x 365
Weight 11.5kg
Outputs: RS 232 C (EDP interface)